Calipers

From Wikipedia, the free encyclopedia
(Redirected from Vernier calipers)
A vernier caliper clamping on an object

Caliper(s) or calliper(s) are an instrument used to measure the linear dimensions of an object or hole; namely, the length, width, thickness, diameter or depth of an object or hole. The word "caliper" comes from a corrupt form of caliber.[1][2][3]

Many types of calipers permit reading out a measurement on a ruled scale, a dial, or an electronic digital display. A common association is to calipers using a sliding vernier scale.

Some calipers can be as simple as a compass with inward or outward-facing points, but with no scale (measurement indication). The tips of the caliper are adjusted to fit across the points to be measured, and then kept at that span while moved to separate measuring device, such as a ruler.

Calipers are used in many fields such as mechanical engineering, metalworking, forestry, woodworking, science and medicine.

Terminology

[edit]

Caliper is the American spelling, while calliper (double "L") is the British spelling.

A single tool might be referred to as a caliper or as calipers — a plural only (plurale tantum) form, like scissors or glasses.

Colloquially, the phrase "pair of verniers" or just "vernier" might refer to a vernier caliper. In loose colloquial usage, these phrases may also refer to other kinds of calipers, although they involve no vernier scale. In machine-shop usage, the term "caliper" is often used in contradistinction to micrometer, even though outside micrometers are technically a form of caliper. In this usage, caliper implies only the form factor of the instrument.

History

[edit]
A bronze sliding pinch scale (collected by the Yangzhou Museum) excavated from the tombs belonging to the Eastern Han dynasty (Gansenshan tombs in Hanjiang District, Yangzhou City)

The earliest caliper has been found in the Greek Giglio wreck near the Italian coast. The ship's find dates to the 6th century BC. The wooden piece already featured a fixed and a movable jaw.[4][5] Although rare finds, calipers remained in use by the Greeks and Romans.[5][6]

A bronze caliper, dating from 9 AD, was used for minute measurements during the Chinese Xin dynasty. The caliper had an inscription stating that it was "made on the gui-you day,[a] the first day[b] of the first month of the first year of Shijianguo.[c]" The calipers included a "slot and pin" and "graduated in inches and tenths of an inch."[7][8]

The modern vernier caliper was invented by Pierre Vernier, as an improvement of the nonius of Pedro Nunes.

Types

[edit]

Inside caliper

[edit]
Two inside calipers

Inside calipers are used to measure the internal size of an object.

  • The upper caliper in the image (on the right) requires manual adjustment prior to fitting. Fine setting of this caliper type is performed by tapping the caliper legs lightly on a handy surface until they will almost pass over the object. A light push against the resistance of the central pivot screw then spreads the legs to the correct dimension and provides the required, consistent feel that ensures a repeatable measurement.
  • The lower caliper in the image has an adjusting screw that permits it to be carefully adjusted without removal of the tool from the workpiece.

Outside caliper

[edit]
Three outside calipers

Outside calipers are used to measure the external size of an object.

The same observations and technique apply to this type of caliper, as for the inside caliper. With some understanding of their limitations and usage, these instruments can provide a high degree of accuracy and repeatability. They are especially useful when measuring over very large distances; consider if the calipers are used to measure a large diameter pipe. A vernier caliper does not have the depth capacity to straddle this large diameter and at the same time reach the outermost points of the pipe's diameter. They are made from high carbon steel.

Divider caliper

[edit]
A pair of dividers

In the metalworking field, a divider caliper, popularly called a compass, is used to mark out locations. The points are sharpened so that they act as scribers; one leg can then be placed in the dimple created by a center or prick punch and the other leg pivoted so that it scribes a line on the workpiece's surface, thus forming an arc or circle.

Their namesake use is in dividing a workpiece of arbitrary width into equal-width sections: by "walking" the tool from one end to the other by pivoting it from one point to the next until reaching the other end, then adjusting the gap between the points until the "walk" ends directly on the end point, equal divisions can be easily marked out without any measuring.

A divider caliper is also used to measure a distance between two points on a map. The two caliper ends are brought to the two points whose distance is being measured. The caliper's opening is then either measured on a separate ruler and then converted to the actual distance, or measured directly on a scale drawn on the map. On a nautical chart the distance is often measured on the latitude scale appearing on the sides of the map: one minute of arc along any great circle, e.g. any longitude meridian, is approximately one nautical mile or 1852 meters.

Dividers are also used in the medical profession. An ECG (also EKG) caliper transfers distance on an electrocardiogram; in conjunction with the appropriate scale, the heart rate can be determined. A pocket caliper version was invented by cardiologist Robert A. Mackin.[9][failed verification]

Oddleg caliper

[edit]
Oddleg calipers

Oddleg calipers, Hermaphrodite calipers, or Oddleg Jennys, as pictured on the left, are generally used to scribe a line at a set distance from the edge of a workpiece. The bent leg is used to run along the workpiece edge while the scriber makes its mark at a predetermined distance, this ensures a line parallel to the edge.

In the diagram at left, the uppermost caliper has a slight shoulder in the bent leg allowing it to sit on the edge more securely. The lower caliper lacks this feature but has a renewable scriber that can be adjusted for wear, as well as being replaced when excessively worn.

Vernier caliper

[edit]
Diagram of a vernier caliper

The labelled parts are

  1. Outside large jaws: used to measure external diameter of an object (like a hollow cylinder) or width of an object (like a rod), diameter of an object (like a sphere).
  2. Inside small jaws: used to measure the internal diameter of an object (like a hollow cylinder or pipe).
  3. Depth probe, or depth rod: used to measure depths of an object (like a small beaker) or a hole.
  4. Main scale (metric): marked every millimeter and helps to measure length correct up to 1 mm.
  5. Main scale (imperial): marked in inches and fractions.
  6. Vernier scale (metric) gives interpolated measurements to 0.1 mm or better.
  7. Vernier scale (imperial) gives interpolated measurements in fractions of an inch.
  8. Retainer: used to block movable part to allow the easy transferring of a measurement.

The calipers in the diagram show a primary reading on the metric scale of about 2.475 cm (2.4 cm read from the main scale plus about 0.075 cm from the vernier scale).

Calipers often have a "zero point error": meaning that the calipers do not read 0.000 cm when the jaws are closed. The zero point error must always be subtracted from the primary reading. Let us assume these calipers have a zero-point error of 0.013 cm. This would give us a length reading of 2.462 cm.

For any measurement, reporting the error on the measurement is also important. Ignoring the possibility of interpolation by eye, both the primary reading and the zero point reading are bounded by plus/minus half the length corresponding to the width of the smallest interval on the vernier scale (0.0025 cm). These are "absolute" errors and absolute errors add, so the length reading is then bounded by plus/minus the length corresponding to the full width of the smallest interval on the vernier scale (0.005 cm). Assuming no systematics affect the measurement (the instrument works perfectly), a complete measurement would then read 2.462 cm ± 0.005 cm.

The vernier, dial, and digital calipers directly read the distance measured with high accuracy and precision. They are functionally identical, with different ways of reading the result. These calipers comprise a calibrated scale with a fixed jaw, and another jaw, with a pointer, that slides along the scale. The distance between the jaws is then read in different ways for the three types.

The simplest method is to read the position of the pointer directly on the scale. When the pointer is between two markings, the user can mentally interpolate to improve the precision of the reading. This would be a simply calibrated caliper, but the addition of a vernier scale allows more accurate interpolation and is the universal practice; this is the vernier caliper.

Vernier, dial, and digital calipers can measure internal dimensions (using the uppermost jaws in the picture at right), external dimensions using the pictured lower jaws, and in many cases depth by the use of a probe that is attached to the movable head and slides along the centre of the body. This probe is slender and can get into deep grooves that may prove difficult for other measuring tools.

The vernier scales may include metric measurements on the lower part of the scale and inch measurements on the upper, or vice versa, in countries that use inches. Vernier calipers commonly used in industry provide a precision to 0.01 mm (10 micrometres), or one thousandth of an inch. They are available in sizes that can measure up to 1828 mm (72 in).[10]

Dial caliper

[edit]
TESA dial caliper

Instead of using a vernier mechanism, which requires some practice to use, the dial caliper reads the final fraction of a millimeter or inch on a simple dial.

In this instrument, a small, precise rack and pinion drives a pointer on a circular dial, allowing direct reading without the need to read a vernier scale. Typically, the pointer rotates once every inch, tenth of an inch, or 1 millimeter. This measurement must be added to the coarse whole inches or centimeters read from the slide. The dial is usually arranged to be rotatable beneath the pointer, allowing for "differential" measurements (the measuring of the difference in size between two objects, or the setting of the dial using a master object and subsequently being able to read directly the plus-or-minus variance in the size of subsequent objects relative to the master object).

The slide of a dial caliper can usually be locked at a setting using a small lever or screw; this allows simple go/no-go checks of part sizes.

Digital caliper

[edit]
Digital caliper

Rather than a rack and pinion, digital calipers use a linear encoder. A liquid-crystal display shows the measurement, which often can switch units between millimeters and fractional or decimal inches. All provide for zeroing the display at any point along the slide, allowing the same sort of differential measurements as with the dial caliper. Digital calipers may contain a "reading hold" feature, allowing the reading of dimensions after use in awkward locations where the display cannot be seen. Like analog calipers, the slide of many digital calipers can be locked using a lever or screw.

Resolution and accuracy

[edit]

Ordinary 150 mm (6 in) digital calipers made of stainless steel have a rated accuracy of 0.02 mm (0.001 in) and a resolution of 0.01 mm (0.0005 in).[11] The same technology is used for longer calipers, but accuracy declines to 0.03 mm (0.001 in) for 100–200 mm (4–8 in) and 0.04 mm (0.0015 in) for 200–300 mm (8–12 in) measurements.[12]

Measurement method

[edit]
US patent filing[13] corresponding to a 1983 German patent DE3340782C2[14] for a "capacitive displacement measuring device with t-shaped scale coatings", whose abstract describes it as follows: a capacitance length and/or angle measuring device has a transducer comprising a stationary part forming a scale and a sensing unit, both provided with capacitive coatings. Changing their overlapping area results in a variable capacitive reactance, which changes the phase position of the electrical signals. A linear function between the mechanical quantity to be measured and the phase portion of the signals can be obtained with an appropriate configuration of the capacitor surfaces.

Many digital calipers contain a capacitive linear encoder. Inexpensive Chinese models have 56 narrow emitter plates and one long receiver plate etched on the sliding display's printed circuit board, which intersect with a repeating pattern of T-shaped plates in the longer "stator" board. The top of the "T" plates intersect with the receiver plate, while the vertical bars of each "T" intersect with the emitter plates. The pitch of each "T" in the stator is slightly less than 8 times the pitch of each emitter plate, so their intersecting capacitive area is not perfectly aligned but rather forms an interference pattern.[15] As the slider moves, these variable capacitances change in a repeating linear fashion. The slider's circuitry counts these repetitions as it slides and achieves finer resolution using linear interpolation of the capacitances.[16] One model sends 8 periodic pulse-width modulation voltage signals (which appear identical but out of phase by 18 of the period),[17] each connected to 7 emitter plates, and the resulting analog signal is read through a single receiver plate.[18] The 1983 German patent DE3340782C2 (see figure) is said to describe the workings.[19]

Other digital calipers contain an inductive linear encoder, which allows robust performance in the presence of contamination such as coolants.[20] Magnetic linear encoders are used in yet other digital calipers.[citation needed]

Serial data output

[edit]

Digital calipers nowadays offer serial data output to expedite repeated measurements, avoid human error, and allow direct data entry into a digital recorder, spreadsheet, statistical process control program, or similar software on a personal computer. Interfacing devices based on RS-232, Universal Serial Bus, or wireless can be built or purchased. The serial digital output varies among manufacturers, but common options are:

  • Mitutoyo's Digimatic interface. This is the dominant name brand interface. Format is 52 bits arranged as 13 nibbles.[21][22][23]
  • Sylvac interface. This is the common protocol for inexpensive, non-name brand, calipers. Format is 24-bit 90 kHz synchronous.[24][25]
  • Starrett[26]
  • Brown & Sharpe[26]
  • Federal
  • Tesa[26]
  • Aldi. Format is 7 BCD digits.[25]
  • Mahr (Digimatic, RS232C, Wireless FM Radio, Infrared and USB)[27]

Micrometer screw caliper

[edit]
Micrometer (screw caliper)

A caliper using a calibrated screw for measurement, rather than a slide, is called an external micrometer caliper gauge, a micrometer caliper or, more often, simply a micrometer. (Sometimes the term caliper, referring to any other type in this article, is held in contradistinction to micrometer.)


Comparison

[edit]

Each of the above types of calipers has its relative merits and faults.

Vernier calipers are rugged and have long-lasting accuracy, are coolant proof, are not affected by magnetic fields, and are largely shockproof. They may have both centimeter and inch scales. However, vernier calipers require good eyesight or a magnifying glass to read and can be difficult to read from a distance or from awkward angles. It is relatively easy to misread the last digit. In production environments, reading vernier calipers all day long is error-prone and is annoying to the workers.

Dial calipers are comparatively easy to read, especially when seeking the exact center by rocking and observing the needle movement. They can be set to 0 at any point for comparisons. They are usually fairly susceptible to shock damage. They are also very prone to getting dirt in the gears, which can cause accuracy problems.

Digital calipers switch easily between centimeter and inch systems. They can be set to zero easily at any point with a full count in either direction and can take measurements even if the display is completely hidden, either by using a "hold" key, or by zeroing the display and closing the jaws, showing the correct measurement, but negative. They can be mechanically and electronically fragile. Most also require batteries and do not resist coolant well. They are also only moderately shockproof and can be vulnerable to dirt.

Calipers may read to a resolution of 0.01 mm or 0.0005 in, but accuracy may not be better than about ±0.02 mm or 0.001 in for 150 mm (6 in) calipers, and worse for longer ones.[28]

Use

[edit]
Using the vernier caliper
A biologist uses calipers to measure the length of a bird's leg

A caliper must be properly applied against the part in order to take the desired measurement. For example, when measuring the thickness of a plate, a vernier caliper must be held at right angles to the piece. Some practice may be needed to measure round or irregular objects correctly.

Accuracy of measurement when using a caliper is highly dependent on the skill of the operator. Regardless of type, a caliper's jaws must be forced into contact with the part being measured. As both part and caliper are always to some extent elastic, the amount of force used affects the indication. A consistent, firm touch is correct. Too much force results in an under indication as part and tool distort; too little force gives insufficient contact and an over indication. This is a greater problem with a caliper incorporating a wheel, which lends mechanical advantage. This is especially the case with digital calipers, calipers out of adjustment, or calipers with a poor quality beam.

Simple calipers are uncalibrated; the measurement taken must be compared against a scale. Whether the scale is part of the caliper or not, all analog calipers—verniers and dials—require good eyesight in order to achieve the highest precision. Digital calipers have an advantage in this area.

Calibrated calipers may be mishandled, leading to loss of zero. When a caliper's jaws are fully closed, it should, of course, indicate zero. If it does not, it must be recalibrated or repaired. A vernier caliper does not easily lose its calibration, but a sharp impact or accidental damage to the measuring surface in the caliper jaw can be significant enough to displace zero.[29] Digital calipers have zero set buttons, for quick recalibration.

Vernier, dial and digital calipers can be used with accessories that extend their usefulness. Examples are a base that extends their usefulness as a depth gauge and a jaw attachment that all allows measuring the center distance between holes. Since the 1970s, a clever modification of the moveable jaw on the back side of any caliper allows for step or depth measurements in addition to external caliper measurements, similarly to a universal micrometer (e.g., Starrett Mul-T-Anvil or Mitutoyo Uni-Mike).

Zero error

[edit]
If when the jaws are closed the reading is 0.10 mm, the zero error is +0.10 mm. The method to use a vernier scale or caliper with zero error is to use the formula "(actual reading) = (main scale) + (vernier scale) − (zero error)", thus the actual reading is 19.00 + 0.54 − (0.10) = 19.44 mm. The resolution of the measurement, based on the width of the smallest sub-interval, is ±0.02 mm.

The method to use a vernier scale or caliper with zero error is to use the formula "actual reading = main scale + vernier scale − (zero error)". Zero error may arise due to knocks that affect the calibration at 0.00 mm when the jaws are perfectly closed or just touching each other. Positive zero error refers to the fact that when the jaws of the vernier caliper are just closed, the reading is a positive reading away from the actual reading of 0.00 mm. If the reading is 0.10 mm, the zero error is referred to as +0.10 mm. Negative zero error refers to the fact that when the jaws of the vernier caliper are just closed, the reading is a negative reading away from the actual reading of 0.00 mm. If the reading is −0.08 mm, the zero error is referred to as −0.08 mm.

Abbe error

[edit]
Comparison of Abbe error for Vernier calipers and a micrometer

Calipers with measurement axes displaced from the object being measured suffer from Abbe error if the jaws are not perpendicular due to manufacturing tolerances. Unlike zero error, the amount of Abbe error depends on the offset.[30]

See also

[edit]

Notes

[edit]
  1. ^ The 10th day of the 60-day cycle.
  2. ^ Lit. "the day of the new moon."
  3. ^ Shijianguo is the first era name of Wang Mang, the first and only emperor of the Xin dynasty, lit. "the beginning of a nation's establishment."

References

[edit]
  1. ^ "caliper". Merriam-Webster. Merriam-Webster Inc. Retrieved 2023-04-22.
  2. ^ "caliper". Collins English Dictionary. HarperCollins Publishers. Retrieved 2023-04-22.
  3. ^ "calipers". Cambridge English Dictionary. Cambridge University Press. Retrieved 2023-04-22.
  4. ^ Mensun Bound: The Giglio wreck: a wreck of the Archaic period (c. 600 BC) off the Tuscany island of Giglio, Hellenic Institute of Marine Archaeology, Athens 1991, pp. 27 and 31 (Fig. 65)
  5. ^ a b Roger B. Ulrich: Roman woodworking, Yale University Press, New Haven, Conn., 2007, ISBN 0-300-10341-7, p.52f.
  6. ^ "hand tool." Encyclopædia Britannica from Encyclopædia Britannica 2006 Ultimate Reference Suite DVD. [Accessed July 29, 2008]
  7. ^ Colin A. Ronan; Joseph Needham (24 June 1994). The Shorter Science and Civilisation in China: 4. Cambridge University Press. p. 36. ISBN 978-0-521-32995-8. adjustable outside caliper gauge... self-dated at AD 9. An abridged version.
  8. ^ "Bronze Caliper of the Wang Mang Regime". Cultural-China.com. Archived from the original on 2014-08-31. Retrieved 2013-11-26.
  9. ^ "Home". mackinmfg.com.
  10. ^ "Slide Calipers" (PDF). Starrett.com. Archived from the original (PDF) on 2010-05-25. Retrieved 2010-06-25.
  11. ^ "Harbor Freight Tools – Quality Tools at Discount Prices Since 1977". www.harborfreight.com. Archived from the original on 2010-04-04. Retrieved 2014-01-07.
  12. ^ "Digital Caliper". MSI-Viking.com. Archived from the original on 2007-07-26. Retrieved 2009-06-12.
  13. ^ US5068653A, Klingler, Otto; Gruhler, Siegfried & Rivinius, Helmut, "Capacitive displacement measuring device with t-shaped scale coatings", issued 1991-11-26 
  14. ^ DE3340782C2, Gruhler, Siegfried Dipl-Ing 7243 Vöhringen; Klingler, Otto Dr-Ing Dr 7238 Oberndorf & Rivinius, Helmut Dipl-Ing 7124 Bönningheim, "Capacitive length and angle measuring device", issued 1985-12-05 
  15. ^ Inside a cheap set of eBay digital calipers, retrieved 2023-10-11
  16. ^ "How does the Electronic Module Work?". Made-in-China.com. 2010-11-15. Archived from the original on 2017-08-12. Retrieved 2023-10-11.
  17. ^ "HP54645D System A.02.07". Archived from the original on 2011-07-19.
  18. ^ Müller, Nick. "How do the Digital Scales work?". Archived from the original on 2014-03-30.
  19. ^ Trebbin, Grant (2014-04-20). "Digital Calliper Teardown and Repair". Archived from the original on 2023-07-05. Retrieved 2023-10-11.
  20. ^ ABS Coolant Proof Caliper; Coolant Proof Micrometer (PDF), Mitutoyo, January 2005, Bulletin No. 1813-293/500, archived from the original (PDF) on 2013-11-03
  21. ^ DRO System: Linear Scales & Counters (PDF), Mitutoyo, n.d., p. 22, Bulletin No. 1715, archived from the original (PDF) on 2013-11-03
  22. ^ Linear Gage (PDF), Mitutoyo, n.d., p. 33, Catalog No. E4174-542/572/575, archived from the original (PDF) on 2013-11-03
  23. ^ Lancaster, Don (February 2000), "Tech Musings" (PDF), Tech Musings, 145
  24. ^ "Chinese Scales". www.shumatech.com.
  25. ^ a b "Protocols of Digital Scales". Yadro.de. Archived from the original on 2013-12-09.
  26. ^ a b c Lancaster, Don (November 1999), "Tech Musings" (PDF), Tech Musings, 142: 142.3
  27. ^ "New Mahr 2015 Catalogue". www.Cutwel.co.uk. Cutwel.
  28. ^ "Accuracy of Calipers". www.tresnainstrument.com.
  29. ^ Mitutoyo. E 12024 Check Points For Measuring Instruments. pp. 2, 3.
  30. ^ Leach, Richard (2014). "Abbe Error/Offset". CIRP Encyclopedia of Production Engineering. pp. 1–4. doi:10.1007/978-3-642-35950-7_16793-1. ISBN 978-3-642-35950-7.