Jump to content

Diaphragm pacing

From Wikipedia, the free encyclopedia
(Redirected from Diaphragmatic pacemaker)

Diaphragm pacing
Electrical stimulation of the phrenic nerve has been known to stimulate respiration for centuries.
Other namesphrenic nerve pacing

Diaphragm pacing (and even earlier as electrophrenic respiration[1][2]) is the rhythmic application of electrical impulses to the diaphragm to provide artificial ventilatory support for respiratory failure or sleep apnea.[3][4] Historically, this has been accomplished through the electrical stimulation of a phrenic nerve by an implanted receiver/electrode,[5] though today an alternative option of attaching percutaneous wires to the diaphragm exists.[6]

History

[edit]

The idea of stimulating the diaphragm through the phrenic nerve was first firmly postulated by German physician Christoph Wilhelm Hufeland, who in 1783 proposed that such a technique could be applied as a treatment for asphyxia.[7][8]: 545–549  French neurologist Duchenne de Boulogne made a similar proposal in 1855, though neither of them tested it.[9] It was not until a year later that Hugo Wilhelm von Ziemssen demonstrated diaphragm pacing on a 27-year-old woman asphyxiated on charcoal fumes by rhythmically faradizing her phrenic nerves, saving her life.[8][10]: 49  Duchenne would later in 1872 declare the technique the "best means of imitating natural respiration".[11] However, advances in mechanical ventilation by the likes of George Poe in the early twentieth century[12] ended up being initially favored over phrenic nerve stimulation.

Harvard researchers Sarnoff et al. revisited diaphragm pacing via the phrenic nerve in 1948, publishing their experimental results on dogs.[1] In a separate publication a few days before, the same group also revealed they had an opportunity to use the technique "on a five-year-old boy with complete respiratory paralysis following rupture of a cerebral aneurysm". Referring to the process as "electrophrenic respiration", Sarnoff was able to artificially respirate the young boy for 52 hours.[13] The technology behind diaphragm pacing was advanced further in 1968 with the publication of doctors John P. Judson and William W. L. Glenn's research on the use of radio-frequency transmission to at whim "adjust the amplitude of stimulation and to control the rate of stimulation externally".[14] Teaming up with Avery Laboratories, Glenn brought his prototype device to the commercial market in the early 1970s.[15] The Avery Breathing Pacemaker received pre-market approval from the FDA in 1987 for "chronic ventilatory support because of upper motor neuron respiratory muscle paralysis" in patients of all ages.[16] In the 1980s, "sequential multipole stimulation" was developed in Tampere, Finland. This technology was commercialized as the Atrostim PNS system and became commercially available in Europe in 1990.[17]

By the early 1990s, long-term evaluations of the technology were being published, with some researchers such as Bach and O'Connor stating that phrenic nerve pacing is a valid option "for the properly screened patient but that expense, failure rate, morbidity, and mortality remain excessive and that alternative methods of ventilatory support should be explored".[18] Others such as Brouillette and Marzocchi suggested that advances in encapsulation and electrode technologies could improve system longevity and reduce damage to diaphragm muscle.[19] Additionally, new surgical techniques such as a thoracoscopic approach began to appear in the late 1990s.[20]

In the mid-2000s, U.S. company Synapse Biomedical began researching a new diaphragm pacing system that would not have to attach to the phrenic nerve but instead depended on "four electrodes implanted in the muscle of the diaphragm to electronically stimulate contraction". The marketed NeuRx device received several FDA approvals under a Humanitarian Device Exemption (HDE), one in 2008 and another in 2011.[21]

Methodology and devices

[edit]

The basic principle behind a diaphragm pacing device (the U.S. Food and Drug Administration identifies the device as a "diaphragmatic/phrenic nerve stimulator"[22]) involves passing an electric current through electrodes that are attached internally. The diaphragm contracts, expanding the chest cavity, and causing air to be sucked into the lungs (inspiration). When not stimulated, the diaphragm relaxes and air moves out of the lungs (expiration).

According to the United States Medicare system, phrenic nerve stimulators are indicated for "selected patients with partial or complete respiratory insufficiency" and "can be effective only if the patient has an intact phrenic nerve and diaphragm".[23] Common patient diagnoses for phrenic nerve pacing include patients with spinal cord injury, central sleep apnea, congenital central hypoventilation syndrome (i.e., Ondine's curse), and diaphragm paralysis.[21][23]

There are currently three commercially distributed diaphragm pacing devices: Synapse Biomedical, Inc.'s NeuRx (US), Avery Biomedical Devices, Inc.'s Mark IV Breathing Pacemaker (US),[21] and Atrotech OY's Atrostim PNS (Finland).[24] The Synapse and Avery devices are distributed worldwide and approved for use in the United States.[21] The Atrotech device is not available in the U.S. As of December 2019, FDA Premarket Approval was given to Avery's Spirit Transmitter Device, replacing the Mark IV transmitter.[25]

In May 2020, Canadian company Lungpacer Medical received approval for emergency use by the USFDA amid the COVID-19 pandemic.[26]

Surgical procedure

[edit]

In the case of the Atrostim and Mark IV devices, several surgical techniques may be used. Surgery is typically performed by placing an electrode around the phrenic nerve, either in the neck (i.e., cervically; an older technique), or in the chest (i.e., thoracically; more modern). This electrode is connected to a radiofrequency receiver which is implanted just under the skin. An external transmitter sends radio signals to the device by an antenna that is worn over the receiver.[27] For the cervical surgical technique, the phrenic nerve is approached via a small (~5 cm) incision slightly above, and midline to, the clavic. The phrenic nerve is then isolated under the scalenus anticus muscle. For the thoracic surgical technique, a small (~5 cm) incisions over the 2nd or 3rd intercostal space. The electrodes are placed around the phrenic nerves alongside the pericardium. The use of a thorascope allows for this technique to be performed in a minimally-invasive manner.[27]

In the case of the NeuRx device, a series of four incisions are made in the abdominal skin. Several tools such as a laparoscope and probe are used to find the best four locations on the diaphragm to attach four electrodes, which have connections outside the body. A fifth electrode is placed just under the skin in the same area. All these connect to the device.[28]

Research

[edit]

Research into the efficacy of diaphragmatic pacing (DP) for ventilatory support has had mixed results, that depend largely on the disease process of the patient population being studied.[29]

Numerous studies have shown that DP can be a very effective tool for patients with spinal cord injuries (SCI).[30][31][32][29] One retrospective long-term multivariate analysis of patients with cervical SCI injuries, found that DP produced better survival rates and updated self-reported quality of life ratings than did the use of mechanical ventilation (MV) .[31]

According to a number of high profile studies, diaphragm pacing should not be used to treat respiratory failure in patients with Amyotrophic Lateral Sclerosis (ALS).[33][34] Studies have demonstrated a significantly higher mortality rate in patients with activated DP implants, than those who received the traditional MV for treatment in later stages of ALS. There have been virtually no clinical trials published on the efficacy of DP in patients with ALS since around 2016, due to the results of the DiPALS study, that showed a mortality rate of 76% of patients with the implant compared to 51% of patients who received MV alone.[34] The USA has allowed the implementation of DP for patients with ALS using a “Compassionate Authorization”, however Europe has stopped all authorization for DP in patients with ALS.[35]

References

[edit]
  1. ^ a b Sarnoff, S.J.; Whittenberger, J.L.; Hardenbergh, E. (1948). "Electrophrenic respiration. Mechanism of the inhibition of spontaneous respiration". American Journal of Physiology. 155 (2): 203–207. doi:10.1152/ajplegacy.1948.155.2.203. PMID 18107083.
  2. ^ Marshall, L.B., ed. (1951). "Electrophrenic Respiration". United States Navy Medical News Letter. 18 (4): 10–12.
  3. ^ Bhimji, S. (16 December 2015). Mosenifar, Z. (ed.). "Overview - Indications and Contraindications". Medscape - Diaphragm Pacing. WebMD LLC. Retrieved 19 February 2016.
  4. ^ Khanna, V.K. (2015). "Chapter 19: Diaphragmatic/Phrenic Nerve Stimulation". Implantable Medical Electronics: Prosthetics, Drug Delivery, and Health Monitoring. Springer International Publishing AG Switzerland. p. 453. ISBN 9783319254487. Retrieved 19 February 2016.
  5. ^ Chen, M.L.; Tablizo, M.A.; Kun, S.; Keens, T.G. (2005). "Diaphragm pacers as a treatment for congenital central hypoventilation syndrome". Expert Review of Medical Devices. 2 (5): 577–585. doi:10.1586/17434440.2.5.577. PMID 16293069. S2CID 12142444.
  6. ^ "Use and Care of the NeuRx Diaphragm Pacing System" (PDF). Synapse Biomedical, Inc. Retrieved 19 February 2016.
  7. ^ Hufeland, C.W. (1783). Usum uis electriciae in asphyxia experimentis illustratum. Dissertatio inauguralis medica sistens.
  8. ^ a b Althaus, Julius (1870). A Treatise on Medical Electricity, Theoretical and Practical: And Its Use in the Treatment of Paralysis, Neuralgia and Other Diseases (2nd ed.). London: Longmans, Green, and Co. p. 676. Retrieved 19 February 2016.
  9. ^ Duchenne, G.B.A. (1855). De l'electrisation localisée et de son application a la physiologie, a la pathologie et a la thérapeutique. Paris: Baillière. Retrieved 19 February 2016.
  10. ^ von Ziemssen, H.W. (1857). Die Electricität in der Medicin Studien. Berlin: Hirschwald. p. 106. Retrieved 19 February 2016.
  11. ^ Duchenne, G.B.A. (1872). De l'électrisation localisée et de son application à la pathologie et à la thérapeutique par courants induits et par courants galvaniques interrompus et continus. Paris: Baillière. Retrieved 19 February 2016.
  12. ^ "Smother Small Dog To See it Revived". The New York Times. 29 May 1908. Retrieved 19 February 2016 – via WikiMedia Commons.
  13. ^ Sarnoff, S.J.; Hardenbergh, E.; Whittenberger, J.L. (1948). "Electrophrenic Respiration". Science. 108 (2809): 482. Bibcode:1948Sci...108..482S. doi:10.1126/science.108.2809.482. PMID 17733839.
  14. ^ Judson, J.P.; Glenn, W.W.L. (1968). "Radio-Frequency electrophrenic respiration: Long-term application to a patient with primary hypoventilation". JAMA. 203 (12): 1033–1037. doi:10.1001/jama.1968.03140120031007. PMID 5694362.
  15. ^ "History of Pacing". Avery Biomedical Devices, Inc. Retrieved 19 February 2016.
  16. ^ Wang, Diep (June 2015). "Diaphragm Pacing without Tracheostomy in Congenital Central Hypoventilation Syndrome Patients". Respiration. 89 (6). Retrieved 2 December 2019.
  17. ^ "Phrenic Nerve Stimulation". Atrotech. Retrieved 2 December 2019.
  18. ^ Bach, J.R.; O'Connor, K. (1991). "Electrophrenic ventilation: A different perspective". The Journal of the American Paraplegia Society. 14 (1): 9–17. doi:10.1080/01952307.1991.11735829. PMID 2022962.
  19. ^ Brouillette, R.T.; Marzocchi, M. (1994). "Diaphragm pacing: clinical and experimental results". Biology of the Neonate. 65 (3–4): 265–271. doi:10.1159/000244063. PMID 8038293.
  20. ^ Shaul, D.B.; Danielson, P.D.; McComb, J.G.; Keens, T.G. (2002). "Thoracoscopic placement of phrenic nerve electrodes for diaphragmatic pacing in children". Journal of Pediatric Surgery. 37 (7): 974–978. doi:10.1053/jpsu.2002.33821. PMID 12077752.
  21. ^ a b c d "Diaphragmatic/Phrenic Nerve Stimulation and Diaphragm Pacing Systems". Policy # MED.00100. Anthem Insurance Companies, Inc. 5 February 2015. Retrieved 19 February 2016.
  22. ^ "PART 882 -- NEUROLOGICAL DEVICES". CFR - Code of Federal Regulations Title 21. U.S. Food and Drug Administration. 21 August 2015. Retrieved 19 February 2016.
  23. ^ a b "Chapter 1, Part 2, Section 160.19: Phrenic Nerve Stimulator" (PDF). Medicare National Coverage Determinations Manual. Centers for Medicare and Medicaid Services. 27 March 2015. Retrieved 19 February 2016.
  24. ^ "Phrenic Nerve Stimulation". Atrotech OY. Retrieved 19 February 2016.
  25. ^ "Premarket Approval". fda.gov. US Food and Drug Administration. Retrieved 2 December 2019.
  26. ^ SFU professor's invention to help recovery of ventilator patients approved for emergency use in U.S. - Meera Bains, Canadian Broadcasting Corporation, 8 May 2020
  27. ^ a b Bhimji, S. (16 December 2015). Mosenifar, Z. (ed.). "Technique - Insertion of Pacemaker". Medscape - Diaphragm Pacing. WebMD LLC. Retrieved 19 February 2016.
  28. ^ "Surgery: What to Expect" (PDF). NeuRx Diaphragm Pacing System Patient/Caregiver Information and Instruction Manual. Synapse Biomedical, Inc. 2011. p. 18. Retrieved 19 February 2016.
  29. ^ a b Woo, A-la; Tchoe, Ha-jin; Shin, Hae-won; Shin, Chae-min; Lim, Chae-Man (2020). "Assisted Breathing with a Diaphragm Pacing System: A Systematic Review". Yonsei Medical Journal. 61 (12): 1024–1033. doi:10.3349/ymj.2020.61.12.1024. ISSN 0513-5796. PMC 7700882. PMID 33251776.
  30. ^ Wijkstra, Peter J.; van der Aa, Hans; Hofker, H. Sijbrand; Curto, Francesco; Giacomini, Matteo; Stagni, Giuliana; Dura Agullo, Maria Asuncion; Curià Casanoves, Francesc Xavier; Benito-Penalva, Jesús; Martinez-Barenys, Carlos; Vidal, Joan (2022). "Diaphragm Pacing in Patients with Spinal Cord Injury: A European Experience". Respiration. 101 (1): 18–24. doi:10.1159/000517401. ISSN 0025-7931. PMC 8820415. PMID 34274930.
  31. ^ a b Romero, F J; Gambarrutta, C; Garcia-Forcada, A; Marín, M A; Diaz de la Lastra, E; Paz, F; Fernandez-Dorado, M T; Mazaira, J (16 July 2021). "Long-term evaluation of phrenic nerve pacing for respiratory failure due to high cervical spinal cord injury". Spinal Cord. 50 (12): 895–898. doi:10.1038/sc.2012.74. ISSN 1362-4393. PMID 22777487.
  32. ^ Posluszny, Joseph A.; Onders, Raymond; Kerwin, Andrew J.; Weinstein, Michael S.; Stein, Deborah M.; Knight, Jennifer; Lottenberg, Lawrence; Cheatham, Michael L.; Khansarinia, Saeid; Dayal, Saraswati; Byers, Patricia M.; Diebel, Lawrence (1 February 2014). "Multicenter review of diaphragm pacing in spinal cord injury: Successful not only in weaning from ventilators but also in bridging to independent respiration". Journal of Trauma and Acute Care Surgery. 76 (2): 303–310. doi:10.1097/TA.0000000000000112. ISSN 2163-0755. PMID 24458038.
  33. ^ Gonzalez-Bermejo, Jésus; Morélot-Panzini, Capucine; Tanguy, Marie-Laure; Meininger, Vincent; Pradat, Pierre-François; Lenglet, Timothée; Bruneteau, Gaëlle; Forestier, Nadine Le; Couratier, Philippe; Guy, Nathalie; Desnuelle, Claude; Prigent, Hélène; Perrin, Christophe; Attali, Valérie; Fargeot, Catherine (11 October 2016). "Early diaphragm pacing in patients with amyotrophic lateral sclerosis (RespiStimALS): a randomised controlled triple-blind trial". The Lancet Neurology. 15 (12): 1217–1227. doi:10.1016/S1474-4422(16)30233-2. PMID 27751553.
  34. ^ a b "Safety and efficacy of diaphragm pacing in patients with respiratory insufficiency due to amyotrophic lateral sclerosis (DiPALS): a multicentre, open-label, randomised controlled trial". The Lancet Neurology. 14 (9): 883–892. 30 July 2015. doi:10.1016/S1474-4422(15)00152-0. hdl:10026.1/3556. PMID 26234554.
  35. ^ Le Pimpec-Barthes, Francoise; Legras, Antoine; Arame, Alex; Pricopi, Ciprian; Boucherie, Jean-Claude; Badia, Alain; Panzini, Capucine Morelot (8 April 2016). "Diaphragm pacing: the state of the art". Journal of Thoracic Disease. 8 (S4): S376–S386. doi:10.21037/jtd.2016.03.97. PMC 4856845. PMID 27195135.

Further reading

[edit]